Common faults and troubleshooting of vacuum hot pressing furnace
Time:2015-07-14 08:51:42
Vacuum hot pressing furnace Common Faults and exclusion methodc2aZhengzhou Protech Technology Co., Ltd
High vacuum pumping on but does not leak (mean value vacuum gauge does not move for a long time) Possible reasons:c2aZhengzhou Protech Technology Co., Ltd
1. High vacuum gauge tube aging c2aZhengzhou Protech Technology Co., Ltd
2. Diffusion pump oil aging Solution: Replace the vacuum gauge; replace the diffusion pump oil. Thermocouple automatic access failure, warning lights alarmc2aZhengzhou Protech Technology Co., Ltd
Possible reasons:c2aZhengzhou Protech Technology Co., Ltd
1. Inner and outer limit failure. c2aZhengzhou Protech Technology Co., Ltd
2. Out of the motor does not turn.c2aZhengzhou Protech Technology Co., Ltd
Solution: Adjust or replace the stopper inside and outside; testing whether the motor is already bad, as bad replacement of the same type of motor. (Note) should be observed before the heating furnace within the limit indicator light should be lit. Low vacuum was not on but does not leak (mean value vacuum gauge does not move for a long time) Possible reasons:c2aZhengzhou Protech Technology Co., Ltd
1. Low vacuum gauge operating current is incorrect. c2aZhengzhou Protech Technology Co., Ltd
2. Mechanical Pump Oil agingc2aZhengzhou Protech Technology Co., Ltd
Solution: Adjust the operating current regulatory vacuum; replace mechanical pump oil. Over-temperature alarm and warning lights onc2aZhengzhou Protech Technology Co., Ltd
Possible reasons:c2aZhengzhou Protech Technology Co., Ltd
1. Over-temperature furnace c2aZhengzhou Protech Technology Co., Ltd
2. Instrument setting incorrect Solution: Press the manual reset over-temperature alarm parameter After a period of time the temperature does not rise up Possible causes: heater resistance value is increased. Solution: According to instructions to transfer large output power meter; severe aging only replacement heater Overcurrent control cabinet panel light is on and alarm Possible causes: Short circuit heater Solution: Open the lid heater observe whether there is a short circuit troubleshooting. Without water light is on and alarm Possible reasons:c2aZhengzhou Protech Technology Co., Ltd
1. Did not open the valve c2aZhengzhou Protech Technology Co., Ltd
2. Low water pressurec2aZhengzhou Protech Technology Co., Ltd
Workaround: open the valve; increase water pressure.